Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Wavetek 2010: (1 year) CAL VER /5520 INSTRUMENT: Wavetek 2020: (1 year) CAL VER /5520 DATE: 12-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 45 NUMBER OF LINES: 195 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke Corp. makes no warranty, expressed or implied, as to the fitness or # useability of this procedure. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: NONE 1.004 DISP Temperature: 18 - 28 degC 1.004 DISP Humidity: <80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {DISPLAY SEGMENT} 1.007 MESS While turning instrument from OFF to ACV, notice that 1.007 MESS ALL display segments light for approx. 2-5 seconds 1.007 MESS 1.008 EVAL Do ALL the display segments light correctly? 2.001 MESS 2.002 HEAD {AC VOLTAGE} 2.003 DISP Connect the 5520A to the UUT as follows: 2.003 DISP [32] 5520A NORMAL HI to UUT V/OHM 2.003 DISP [32] 5520A NORMAL LO to UUT COM 2.004 DISP Set the Function Switch to ACV. 2.005 HEAD AC VOLTAGE: {100mV Range} 2.006 5520 100 90.00mV 2% 0.08/ 400H SI 2W 3.001 HEAD AC VOLTAGE: {1000mV Range} 3.002 5520 1000 900.0mV 2% 0.08/ 60H SI 2W 4.001 5520 1000 900.0mV 2% 0.08/ 2kH SI 2W 5.001 HEAD AC VOLTAGE: {10V Range} 5.002 5520 10 9.000V 2% 0.08/ 400H SI 2W 6.001 HEAD AC VOLTAGE: {100V Range} 6.002 5520 100 90.00V 2% 0.08/ 400H SI 2W 7.001 HEAD AC VOLTAGE: {1000V Range} 7.002 5520 1000 700.0V 2% 0.08/ 400H SI 2W 8.001 HEAD {DC VOLTAGE} 8.002 DISP Set the Function Switch to DCV. 8.003 HEAD DC VOLTAGE: {100mV Range} 8.004 5520 100 90.00mV 0.25% 0.02/ 2W 9.001 HEAD DC VOLTAGE: {1000mV Range} 9.002 5520 1000 900.0mV 0.25% 0.02/ 2W 10.001 HEAD DC VOLTAGE: {10V Range} 10.002 5520 10 9.000V 0.25% 0.02/ 2W 11.001 HEAD DC VOLTAGE: {100V Range} 11.002 5520 100 90.00V 0.25% 0.02/ 2W 12.001 HEAD DC VOLTAGE: {1000V Range} 12.002 5520 1000 900.0V 0.25% 0.02/ 2W 13.001 HEAD {DIODE TEST} 13.002 DISP Set the Function Switch to DT. 13.003 5520 0.5000V 1% 0.0005U 2W 14.001 HEAD {RESISTANCE} 14.002 HEAD RESISTANCE: {20 MOhm Range} 14.003 5520 20 20.00MZ 2% 0.04U 2W 15.001 HEAD RESISTANCE: {10 MOhm Range} 15.002 5520 10 9.000MZ 1.5% 0.04/ 2W 16.001 HEAD RESISTANCE: {1000 kOhm Range} 16.002 5520 1000 900.0kZ 0.5% 0.04/ 2W 17.001 HEAD RESISTANCE: {100 kOhm Range} 17.002 DISP Connect the 5520A to the UUT as follows: 17.002 DISP [32] 5520A AUX HI to UUT V/OHM 17.002 DISP [32] 5520A AUX LO to UUT COM 17.002 DISP [32] 5520A NORMAL HI to UUT V/OHM 17.002 DISP [32] 5520A NORMAL LO to UUT COM 17.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 17.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 17.003 DISP Set the Function Switch to OHMS. 17.004 5520 100 90.00kZ 0.5% 0.04/ CW 18.001 HEAD RESISTANCE: {10 kOhm Range} 18.002 5520 10 9.000kZ 0.5% 0.04/ CW 19.001 HEAD RESISTANCE: {1000 Ohm Range} 19.002 5520 1000 900.0Z 0.5% 0.04/ CW 20.001 HEAD RESISTANCE: {100 Ohm Range} 20.002 5520 100 90.00Z 0.7% 0.04/ CW 21.001 HEAD {CAPACITANCE} 21.002 DISP Set the Function Switch to CAP. 21.003 HEAD CAPACITANCE: {2uF Range} 21.004 5520 2 1.00uF 2% 0.02U CW 22.001 HEAD CAPACITANCE: {20uF Range} 22.002 5520 20 10.0uF 2% 0.2U CW 23.001 HEAD CAPACITANCE: {200uF Range} 23.002 5520 200 100uF 2% 3U CW 24.001 HEAD CAPACITANCE: {2mF Range} 24.002 5520 2.00 1.00mF 2% 0.08U CW 25.001 DISP Connect the 5520A to the UUT as follows: 25.001 DISP [32] 5520A NORMAL HI to UUT V-OHM 25.001 DISP [32] 5520A NORMAL LO to UUT COM 25.002 HEAD CAPACITANCE: {20nF Range} 25.003 5520 20 10.0nF 2% 0.3U 2W 26.001 HEAD CAPACITANCE: {200nF Range} 26.002 5520 200 100nF 2% 3U 2W 27.001 HEAD {FREQUENCY} 27.002 DISP Set the Function Switch to HZ 27.003 HEAD FREQUENCY: {20kHz Range} 27.004 5520 20 10.000kH 0.1% 0.02/ 100mV SI 2W 28.001 HEAD FREQUENCY: {200kHz Range} 28.002 5520 200 100.00kH 0.1% 0.02/ 100mV SI 2W 29.001 HEAD FREQUENCY: {2MHz Range} 29.002 5520 2 1.0000MH 0.1% 0.02/ 1V SI 2W 30.001 HEAD {PULSE MODE (Duty Cycle)} 30.002 DISP Set the Function Switch to DUTY CYCLE. 30.003 M5520 1Vpp 30.004 ASK- U 30.005 5520 25.0pct 200H SQ S 2W 30.006 ASK+ U 30.007 ACC 25.0pct 0.84U 30.008 MEMI Enter UUT reading in percent. 30.009 MEME 30.010 MEMC pct 2% 0.1U 31.001 ASK- U 31.002 5520 50.0pct 10kH SQ S 2W 31.003 ASK+ U 31.004 ACC 50.0pct 2.8U 31.005 MEMI Enter UUT reading in percent. 31.006 MEME 31.007 MEMC pct 1% 0.1U 32.001 ASK- U 32.002 5520 75.0pct 30kH SQ S 2W 32.003 ASK+ U 32.004 ACC 75.0pct 1.2U 32.005 MEMI Enter UUT reading in percent. 32.006 MEME 32.007 MEMC pct 5% 0.1U 33.001 M5520 * 33.002 HEAD {LOW AC CURRENT} 33.003 DISP Set the Function Switch to ACA. 33.004 DISP Connect the 5520A to the UUT as follows: 33.004 DISP [32] 5520A AUX HI to UUT 100mA 33.004 DISP [32] 5520A AUX LO to UUT COM 33.005 HEAD LOW AC CURRENT: {1000uA Range} 33.006 5520 1000 900.0uA 1.7% 0.08/ 400H SI 2W 34.001 HEAD LOW AC CURRENT: {10mA Range} 34.002 5520 10 9.000mA 1.7% 0.08/ 400H SI 2W 35.001 HEAD LOW AC CURRENT: {100mA Range} 35.002 5520 100 90.00mA 1.7% 0.05/ 400H SI 2W 36.001 HEAD {LOW DC CURRENT} 36.002 DISP Set the Function Switch to DCA. 36.003 HEAD LOW DC CURRENT: {1000uA Range} 36.004 5520 1000 900.0uA 1% 0.02/ 2W 37.001 HEAD LOW DC CURRENT: {10mA Range} 37.002 5520 10 9.000mA 0.5% 0.01/ 2W 38.001 HEAD LOW DC CURRENT: {100mA Range} 38.002 5520 100 90.00mA 0.5% 0.01/ 2W 39.001 HEAD {HIGH DC CURRENT } 39.002 DISP Connect the 5520A to the UUT as follows: 39.002 DISP [32] 5520A AUX HI to UUT 10A 39.002 DISP [32] 5520A AUX LO to UUT COM 39.003 HEAD HIGH DC CURRENT: {1000mA Range} 39.004 5520 1000 900.0mA 1% 0.2/ 2W 40.001 HEAD {HIGH AC CURRENT } 40.002 DISP Set the Function Switch to ACA. 40.003 HEAD HIGH AC CURRENT: {1000mA Range} 40.004 5520 1000 900.0mA 1.7% 0.5U 400H SI 2W 41.001 HEAD {HIGH DC CURRENT } 41.002 DISP Connect the 5520A to the UUT as follows: 41.002 DISP [32] 5520A 20A to UUT 10A 41.002 DISP [32] 5520A AUX LO to UUT COM 41.003 DISP Set the Function Switch to DCA. 41.004 HEAD HIGH DC CURRENT: {10A Range} 41.005 5520 10 9.000A 1% 0.01U HC 2W 42.001 HEAD HIGH DC CURRENT: {20A Range} 42.002 5520 20 11.00A 1% 0.01U HC 2W 43.001 HEAD {HIGH AC CURRENT} 43.002 DISP Set the Function Switch to ACA. 43.003 HEAD HIGH AC CURRENT: {10A Range} 43.004 5520 10 9.000A 1.7% 0.003U 400H SI HC 2W 44.001 HEAD HIGH AC CURRENT: {20A Range} 44.002 5520 20 11.00A 1.7% 0.03U 400H SI HC 2W 45.001 END